Defibrillator

MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT

The information is published following the requirements of the Resolution of the Cabinet of Ministers of Ukraine No. 710 dated 11.10.2016

Defibrillator with non-invasive cardiac stimulation (EZS code DK 021:2015:33180000-5 Equipment for maintaining physiological functions of the body (33182100-0) (code NK 024:2019:37806 - Manual external defibrillator))

 

Requirements

Compliance

 

Must be designed for electrical impulse therapy at the stage of cardiac resuscitation, including transthoracic defibrillation with automatic surface defibrillation, synchronized cardioversion, transcutaneous pacing, as well as for ECG monitoring

 

Must be able to be used in adults and children

 

Availability of a color liquid crystal TFT monitor

 

Availability of functional check: charging, battery status, recording, defibrillation curve, alarm, and speech information

 

Availability of operating modes:

  • Asynchronous defibrillation
  • Synchronous defibrillation (cardioversion)
  • Automatic surface defibrillation
  • Non-invasive cardiac pacing

 

The device must have a biphasic pulse shape

 

The ability to provide a pulse shape of the second phase of defibrillation, which is unchanged and does not depend on the contact impedance

 

Availability of voice instructions during the preparation and conduct of defibrillation

 

Indication on the handle of the quality of contact of the electrodes with the patient's body

 

The device must be equipped with an ECG monitoring mode

 

Availability of manual and automatic ECG recording modes

 

Must be able to operate on mains and rechargeable battery

 

The number of defibrillations, when powered by a battery with an energy of at least 270 J without recharging the defibrillator, should be at least 70 times

 

A charging time of 200 J should be no more than 4 s, charging time of 270 J should be no more than 5 s

 

The presence of defibrillation modes is not worse:

  • manual mode
  • manual synchronization mode with the R wave on the ECG
  • AED mode
  • AED mode (children's)

 

The maximum output energy of the charge is not more than 270 J

 

Availability of at least 3 alarm priorities: critical, warning, recommending

 

The presence of an alarm indicator light that changes color depending on the priority

 

Availability of an SD card slot for data storage and software installation

 

Battery charge indicator on the display

 

Battery operation in monitoring mode for at least 180 minutes

 

Availability of visual and audible alarms

 

Availability of built-in thermal printer

 

Availability of a memory card slot

 

 

The delivery set must include (provide a warranty letter):

defibrillator with reusable electrodes - 1 pc;

set of disposable electrodes and adapter for non-invasive cardiac pacing - 1 pc;

ECG cable for 3 leads - 1 pc;

rechargeable battery - 1 pc;

power cable - 1 pc;  

operation manual - 1 pc.