Операційний стіл

                                          MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT

Operating table (EZS code DK 021:2015:33160000-9 Equipment for operating units (331621000-4) (code NK 024:2019:35379 - Universal operating table, electro-hydraulic)

Name of technical or another requirement Compliance Yes/No

1

The table is designed for surgical operations in operating rooms. Together with additional devices it can be used during various surgical operations

 

2

The table panel should consist of 6 sections: head, back (2 parts), central (hip), leg (right and left)

 

3

The table panel must be X-ray transparent

 

4

The outer parts of the table and additional devices must be made of stainless steel

 

5

The base of the table must be made of ABS plastic and must be mobile: 2 swivel and 2 directional wheels, with a diameter of at least 100 mm

 

6

The table must have an electro-hydraulic drive for controlling the table positions: height adjustment, back section, Trendelenburg / anti-Trendelenburg, side tilts, longitudinal movement of the panel, central locking

 

7

Head and foot sections should be mechanically adjustable

 

8

The presence of a renal surgical position, adjustment of at least 130º - carried out mechanically

 

9

The function of longitudinal movement of the table panel, adjustment of at least 300 mm - using an electro-hydraulic drive

 

10

Length of the table panel, not less than 2100 mm

 

11

Table panel width, not less than 550 mm

 

12

Height adjustment of the table panel (min/max) 595 mm / 1035 mm (±5%) - using the electro-hydraulic drive

 

13

Adjustment of the longitudinal inclination of the panel (Trendelenburg / Anti-Trendelenburg), not less than -30º / +30º - using an electro-hydraulic drive

 

14

Adjustment of the lateral tilt of the table panel, not less than +20º/-20° - using the electro-hydraulic drive

 

15

Adjustment of the back section tilt (up / down), not less than +80º / -45° - using the electro-hydraulic drive

 

16

Tilt adjustment of the head section (up / down), not less than +30º / -90° - with gas springs

 

17

Adjustment of inclinations of the leg section (up / down), not less than +30º / -80° - with gas springs

 

18

The leg sections must be able to turn in the horizontal plane (left / right) in the range of at least 0º - 80°

 

19

Flex / Reflex position adjustment: not less than 225° / 100° - using the electro-hydraulic drive

 

20

Availability of side rails for attaching removable devices

 

21

The remote control must have indicators of the switched-on state of the remote control, battery charge level, and operating status control

 

22

The mattresses of the table panel must be separate, correspond to the sections of the table panel, and be made of polyurethane with antibacterial coating using seamless technology, thickness not less than 50 mm

 

23

The presence of a central braking system, which is regulated by an electro-hydraulic drive

 

24

Red emergency stop button at the base of the table

 

25

Safe working load, not less than 250 kg

 

26

The table shall be supplied with the following additional accessories:

- anesthesia shield (1 pc.);

- hand support with straps (2 pcs.);

- tripod for long-term infusions (1 pc.)

- belt for fixing the body (1 pcs.)

- shoulder supports (2 pcs.);

- side stops (2 pcs.);

- strap for fixing the hip (2 pcs.);

- footrests according to Goepel (2 pcs.)

 