MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT
Automated biochemical analyzer (EHS code DK 021:2015:38430000-8 Detectors and analyzers (38432000-2) (code NK 024:2019:56669 - Biochemical automatic analyzer of metabolic profile IVD, stationary))
№ |
Medical and technical requirements | Compliance with the requirements (indicate yes/no) of the user manual |
1 |
Automatic biochemical analyzer, with free access to patients and tests (Random Access) and with discrete access - test-by-test system (Bench Top) - 1 set |
|
2 |
Fully automatic measuring system |
|
3 |
Measured parameters - designed to determine the biochemical parameters of blood - substrates (glucose, urea, cholesterol, triglycerides, protein, uric acid, and others), enzymes (alanine aminotransferase, aspartate aminotransferase, alkaline phosphatase, alpha-amylase, and others) and ions (iron, copper, calcium, magnesium, phosphorus, and others), specific proteins (transferrin, immunoglobulins, C-reactive protein, and others). |
|
4 |
Sample type: serum, plasma, cerebrospinal fluid, urine |
|
5 |
Productivity up to 200 tests/hour |
|
6 |
Number of tests on board (up to 50) |
|
7 |
Number of positions for samples - 50 positions |
|
8 |
Number of positions for reagents - 50 positions |
|
9 |
The minimum volume of the reaction mixture - is 180 µl (180-550 µl) |
|
10 |
Sample dosing volume 2-70 µl (in 0.1 µl increments) |
|
11 |
Reagent dosing volume: Reagent 1: 50-300 µl (in 1 µl increments); Reagent 2: 10-200 µl (in 1 µl increments) |
|
12 |
Incubation time - 1 stage of analysis (using R1) max 648 seconds; For 2 stages of analysis (using R1 and R2) 1st reaction 306 seconds + 2nd reaction 324 seconds |
|
13 |
Analyzer operating cycle - no more than 18 seconds |
|
14 |
Automatic dilution of samples when exceeding the permissible range during measurements. Ability to increase or decrease the dosed volume after a restart. Ability to perform dilution from 2 to 150 times |
|
15 |
Reaction carousel: 45 reusable quartz glass cuvettes, temperature 37±0.2°C |
|
16 |
The optical system of 8 light filters: 340, 405, 505, 546, 578, 600, 660, 700 nm |
|
17 |
Analysis methods: endpoint with reagent and sample blank, kinetics, two-point kinetics, direct potentiometry, immunoturbidimetric, differential, mono- and bichromatic methods. |
|
18 |
Measurement principle: absorption photometry, immunoturbidimetry, potentiometry |
|
19 |
Urgent analyses are processed without interrupting the planned analysis |
|
20 |
Cooled reagent rotor on board |
|
21 |
Calibration type: by standard (calibrator), by calculation factor, multicalibration: linear, nonlinear |
|
22 |
Primary rinsing, automatic rinsing, probe washing, cuvette rinsing |
|
23 |
Option to save the history of auto wash |
|
24 |
Automatic preliminary and final sample dilution |
|
25 |
Sample dispenser with built-in liquid level detector and breakage protection |
|
26 |
Reagent dispenser with built-in liquid level detector and breakage protection |
|
27 |
Cuvette washing: Washing station on board, with continuous washing and drying cycle |
|
28 |
Water consumption: up to 7.5 liters per hour |
|
29 |
Photometric range 0-3 opt. units |
|
30 |
Light source - halogen lamp 12V/20W |
|
31 |
Detector - silicone photodiode |
|
32 |
Communication with external computer - USB interface for Pentium IV and above |
|
33 |
Dimensions - 810 mm (H) x 700 mm (L) x 600 mm (W) |
|
34 |
Weight - 120 kg |
|
35 |
Power supply Alternating current 220 V±10, 50±1 Hz |
|
36 |
Quality control - graphing and verification of results according to Westgard rules |
|
37 |
The complete set:
|
|