Anesthesiological console

MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT

 

Anesthesiology console (EHS code DK 021:2015:33170000-2 Equipment for anesthesia and resuscitation (331721000-7) (NK code 024:2019:37342 Suspended console for infusion system, with ceiling mount))

 

 

Requirements

Compliance

(yes/no)

1

The ceiling console for medical gas and power supply shall consist of the following components: ceiling mounting system, ceiling connecting plates, gas valves, ceiling cover, support brackets, the vertical module with accessories

 

2

The ceiling mounting plate must be equipped with a strip for electricity and gas supply. The electrical connection strip shall have a sufficient number of terminals for connecting/disconnecting the power supply circuits. The gas connection bar must be equipped with service valves to ensure that the gas supply to the column is shut off in case of maintenance or breakdown

 

3

The presence of a round ceiling cap must be made of smooth plastic. The design should be removable for easy access to communications during maintenance

 

4

The console must have a ceiling mounting system on one rotating arm

 

5

The length of the console arm is at least 900 mm

 

6

Rotation range of the console arm - not less than 330⁰

 

7

The rod connection must be rectangular or oval, smooth, with rounded edges, made of high-strength alloy, and easy to clean

 

8

Availability of brakes on the column bar - pneumatic and friction. The design of the brakes should provide a stable stop of the column

 

9

The presence of brake release button should be located in an ergonomic, horizontally oriented one-piece two-handed handle mounted on the front of the shelf. The brake action points shall be color coded

 

10

Vertical rectangular module with parallel mounted distribution panels: height - not more than 700 mm, width - not more than 280 mm, thickness - not more than 210 mm

 

11

The presence of vertical rails must be built into the front, side, and rear walls of the main unit for the installation of shelves, additional rails, and other accessories.

The width of the guides is not more than 10 mm

 

12

Availability of distribution panels - at least 4, on which sockets can be placed

 

13

Availability of standard DIN sockets for medical gas built into the rear panel:

  • oxygen (O2) - 2 pcs.
  • air (AIR5) - 1 pc.
  • vacuum (VAC) - 3 pcs.
  • anesthetic gas outlet (AGSS) - 1 pc.

 

14

Medical gas outlet points must have CE marking, permanent marking and color designation of the gas type in accordance with PN-ISO 32

 

15

Availability of electrical wiring:

220 V electrical sockets with a protective cover and a light indicator or LED signal of the presence of voltage in the sockets, at least 8 pcs.
equipotential socket or terminal block - 2 pcs.

 

16

Availability of telecommunication sockets located in the side panels of the head unit, RJ45 connector - 1 pc.

 

17

Availability of socket with HDMI connector - 1 pc

 

18

The electronic working module for medical gases must comply with the requirements of PN EN 60601-1-2:2007 and PN-EN 60601-1-8:2007, which visually signals the correct/incorrect gas pressure. The module shall be built into the rear panel

 

19

Availability of no more than 3 shelves for medical instruments with dimensions not exceeding 450x500x35 mm +/- 5mm with protective soft plastic buffers to prevent damage to medical equipment and personnel. The upper part of the shelves shall be made of light metal with an antibacterial coating.

Height adjustment - stepless.

Shelf load capacity - no more than 50 kg

 

20

The presence under the lower shelf of a drawer with an internal depth of not more than 100 mm

 

21

Availability of a rack for long-term infusions with 4 hooks mounted on a vertical rail.

 

22

The carrying capacity of the console should be no more than 150 kg

 

 