MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT
Operating table (code according to EHS DK 021: 2015: 33190000-8 - Medical equipment and medical devices, miscellaneous (code 024:2019:36867 - Operating table, universal, mobile))
№ |
Name | Quantity |
1 |
Operating table |
1 pc. |
No. of pos. | Technical indicators (characteristics) | Compliance with the function, or the value of the parameter |
Compliance yes/no, regarding the page(s) of the manufacturer's technical document |
|
The operating table should be designed to accommodate and ensure the optimal position of the patient during surgery | compliance |
|
|
The operating table shall have a hydro-mechanical drive | compliance |
|
|
The hydro-mechanical drive of the table must provide: lifting and lowering of the panel, Trendelenburg and anti-Trendelenburg positions, side tilts of the panel, tilts of the back section, the tilt of the head section, tilt of the leg section | compliance |
|
|
The table shall be equipped with wheels for easy movement in the operating room. The wheels shall be built into the base of the table and covered with stainless steel covers | compliance |
|
|
The table shall have a locking system for fixing the table in the operating room | compliance |
|
|
The base of the table should not exceed the dimensions of the panel, should ensure the stability of the table, and not interfere with the surgeon and his assistants | compliance |
|
|
The base of the table must be protected by stainless steel covers | compliance |
|
|
Metal parts of the table panel shall be made of stainless steel | compliance |
|
|
The table must have unified stainless steel rails for attaching removable devices | compliance |
|
|
The table panel should consist of sections: head, back, central (hip), leg | compliance |
|
|
The leg section shall consist of two parts | compliance |
|
|
The central (hip) section shall have a cutout for gynecological/urological operations | compliance |
|
|
The table panel shall be radiotransparent | compliance |
|
|
The design of the table must provide the possibility of using an X-ray machine and a C-shaped X-ray system during the operation | compliance |
|
|
Table loading capacity, not less than |
170 kg |
|
|
Table weight, not more than |
180 kg |
|
|
Adjustment of leg sections | mechanical |
|
|
Adjustment of the head section | mechanical |
|
|
Height of the table: in the lowest position, not more than in the extreme upper position, not less than |
760 mm 1010 mm |
|
|
Length of the table panel, not less than |
2010 mm |
|
|
Width of the table panel |
500 mm |
|
|
Trendelenburg / anti-Trendelenburg position, not less than |
+25/ -25 |
|
|
A tilt of the back section, up / down, not less than |
+75/ -15 |
|
|
Side tilt of the panel left / right, not less than |
+20/ -20 |
|
|
The tilt of the leg section, up / down, not less than |
+15/ -90 |
|
|
The tilt of the head section, up / down, not less than |
+60/ -60 |
|
|
The leg sections should be separated in the horizontal plane by an angle of at least |
180 |
|
|
Mattress thickness, not more than |
50 мм |
|
|
The presence of a built-in kidney roller with a lifting height, not less than |
120 мм |
|
|
Availability of two additional sections for positioning and supporting the patient during operations on the back, shoulders, and chest cavity | compliance |
|
|
Availability of a special section for positioning and supporting the patient during stomach operations, which is attached to the table panel | compliance |
|
|
Removable devices supplied with the table: |
|
|
|
Set of accessories for general surgery, gynecology/urology |
|
|
32 |
Nail holder |
2 pcs |
|
33 |
Hand holder |
2 pcs |
|
34 |
Side support for the body |
2 pcs |
|
35 |
Anesthesiological screen |
1 pc |
|
36 |
Shoulder support |
2 pcs |
|
(с) 2023