                                                                                                              MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT

 

Sterilization device  (EHS code DK 021:2015:33190000-8 Medical equipment and medical devices, miscellaneous (33191100-6) (code of NC 024:2019:38671 Steam sterilizer))

 

 

Characteristics Meaning

Conformity

(yes/no)

regarding the manufacturer's technical documentation pages

  1.  
Sterilization devise. Availability 1 pc

 

1.1.

Motorized closing of the sterilization chamber

Conformity

 

1.2.

The material of the chamber must be made of stainless steel

Conformity

 

1.3. Chamber volume in liters Not less than 23 liters.

 

1.4. The weight of the device is not more than 65 kg Compliance

 

1.5. Number of trays for tools Not less than 4

 

1.6. Tray for sterilization of instruments Not less than 400x150x10

 

1.7. The sterilizer must have a display showing the temperature and pressure in the chamber, the type of cycle, and the sterilization time Compliance

 

1.8. Built-in tank for clean water Compliance

 

1.9. The capacity of the tank for clean water Not less than 4 liters

 

1.10. Built-in tank for dirty water Compliance

 

1.11. The capacity of the tank for dirty water Not less than 4 liters

 

1.12. Programmable delay of sterilization start Compliance

 

1.13. Preheating function of the device to reduce cycle time Compliance

 

1.14. Additional drying mode Availability

 

1.15. Safety valve and electronic pressure sensor Availability

 

1.16. Built-in pump for filling with water Availability

 

1.17. Water quality control system Compliance

 

1.18. Available cycle types - B, S, N Compliance

 

1.19. Ability to sterilize solid materials Compliance

 

1.20. Ability to sterilize packaged materials Compliance

 

1.21. Ability to sterilize unpackaged materials Compliance

 

1.22. Ability to sterilize porous materials Compliance

 

1.23. Availability of a built-in printer Compliance

 

1.24. Availability of test modes Not less than two

 

1.25. The type of vacuum creation is not worse than Fractional type

 

1.26. Number of sterilization modes Not less than ten

 

1.27.

Parameters of steam sterilization mode:

- maximum working pressure

Not less than 2.5 bar

 

1.28. Parameters of the steam sterilization mode: Availability of sterilization modes within 121°-134°C

 

1.29.

Parameters of the steam sterilization mode:

- Minimum sterilization exposure time in class B sterilization mode

No more than 45 minutes

 

1.30.

Parameters of the steam sterilization mode:

- minimum sterilization exposure time in class S sterilization mode

No more than 35 min

 

1.31. Supply voltage 230 V (+/- 10%) Compliance

 

 