MEDICAL AND TECHNICAL REQUIREMENTS FOR THE SUBJECT OF PROCUREMENT
Sterilization device (EHS code DK 021:2015:33190000-8 Medical equipment and medical devices, miscellaneous (33191100-6) (code of NC 024:2019:38671 Steam sterilizer))
№ |
Characteristics | Meaning |
Conformity (yes/no) regarding the manufacturer's technical documentation pages |
|
Sterilization devise. | Availability 1 pc |
|
1.1. |
Motorized closing of the sterilization chamber |
Conformity |
|
1.2. |
The material of the chamber must be made of stainless steel |
Conformity |
|
1.3. | Chamber volume in liters | Not less than 23 liters. |
|
1.4. | The weight of the device is not more than 65 kg | Compliance |
|
1.5. | Number of trays for tools | Not less than 4 |
|
1.6. | Tray for sterilization of instruments | Not less than 400x150x10 |
|
1.7. | The sterilizer must have a display showing the temperature and pressure in the chamber, the type of cycle, and the sterilization time | Compliance |
|
1.8. | Built-in tank for clean water | Compliance |
|
1.9. | The capacity of the tank for clean water | Not less than 4 liters |
|
1.10. | Built-in tank for dirty water | Compliance |
|
1.11. | The capacity of the tank for dirty water | Not less than 4 liters |
|
1.12. | Programmable delay of sterilization start | Compliance |
|
1.13. | Preheating function of the device to reduce cycle time | Compliance |
|
1.14. | Additional drying mode | Availability |
|
1.15. | Safety valve and electronic pressure sensor | Availability |
|
1.16. | Built-in pump for filling with water | Availability |
|
1.17. | Water quality control system | Compliance |
|
1.18. | Available cycle types - B, S, N | Compliance |
|
1.19. | Ability to sterilize solid materials | Compliance |
|
1.20. | Ability to sterilize packaged materials | Compliance |
|
1.21. | Ability to sterilize unpackaged materials | Compliance |
|
1.22. | Ability to sterilize porous materials | Compliance |
|
1.23. | Availability of a built-in printer | Compliance |
|
1.24. | Availability of test modes | Not less than two |
|
1.25. | The type of vacuum creation is not worse than | Fractional type |
|
1.26. | Number of sterilization modes | Not less than ten |
|
1.27. |
Parameters of steam sterilization mode: - maximum working pressure |
Not less than 2.5 bar |
|
1.28. | Parameters of the steam sterilization mode: | Availability of sterilization modes within 121°-134°C |
|
1.29. |
Parameters of the steam sterilization mode: - Minimum sterilization exposure time in class B sterilization mode |
No more than 45 minutes |
|
1.30. |
Parameters of the steam sterilization mode: - minimum sterilization exposure time in class S sterilization mode |
No more than 35 min |
|
1.31. | Supply voltage 230 V (+/- 10%) | Compliance |
|